HOME CHOICEPRO
Report
- Report Number
- 1423500-2011-03295
- Event Type
- Malfunction
- Date Received
- March 17, 2011
- Date of Event
- January 25, 2011
- Report Date
- February 22, 2011
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K053512
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). BAXTER CONTACTED THE NURSE TO NOTIFY OF THIS INCIDENT OF IIPV THAT WAS FOUND DURING THE DEVICE LOG REVIEW. THE HOMECHOICE (HC) DEVICE WAS RECEIVED OPERATIONAL AND IN GOOD CONDITION FOR EVALUATION AT THE PRODUCT ANALYSIS LAB (PAL). A REVIEW OF THE DEVICE LOGS CONFIRMED THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT. THE CAUSE FOR THE IIPV FOUND IN THE LOGS WAS DETERMINED TO BE INSUFFICIENT DRAIN DUE TO USE ERROR; THE CLINICIAN INAPPROPRIATELY SET MINIMUM DRAIN VOLUME PERCENTAGE SETTING TOO LOW. A LABELING REVIEW OF THE HOMECHOICE APD SYSTEMS PATIENT AT-HOME GUIDE ISSUED WAS FOUND TO BE SUFFICIENT. THE DEVICE'S SERVICE HISTORY RECORD WAS REVIEWED AND NO ISSUES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE IIPV. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
(B)(4). THE DEVICE HAS BEEN RECEIVED, AND THE EVALUATION IS IN PROCESS. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
DURING INITIAL ASSESSMENT OF A RETURNED HOMECHOICE MACHINE, A BAXTER TECHNICIAN FOUND AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) THAT WAS IDENTIFIED IN PATIENT THERAPY LOG ON (B)(6) 2011. THE DRAIN VOLUME WAS 3928 ML DURING CYCLE 5. THE FILL VOLUME WAS 2000 ML. THIS EVENT MEETS OVERFILL CRITERIA. THERE WAS PATIENT INVOLVEMENT, BUT NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOME CHOICEPRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |