FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2022402 · Received February 16, 2011

Report

Report Number
1824206-2011-00903
Event Type
Malfunction
Date Received
February 16, 2011
Date of Event
February 4, 2011
Report Date
February 4, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN FOUND THE BED IN THE BEDSHOP. FOUND THE HEAD UP SOLENOID WAS STUCK IN CLOSED POSITION. REPLACED VALVE #6543035 TO RESOLVE THIS ISSUE.

Description of Event or Problem · 1

INFORMATION ALLEGED THAT THE HEAD OF THE BED WOULD NOT RAISE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. P1900AF1619

Patients

Seq Age Sex Outcome Treatment
1