FDA Adverse Event Malfunction Summary report: N

MINMED QUICK SET

MDR report key: 20221998 · Received September 16, 2024

Report

Report Number
3003442380-2024-25566
Event Type
Malfunction
Date Received
September 16, 2024
Date of Event
August 16, 2024
Report Date
October 21, 2024
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
PMA / PMN Number
K160648
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: THE COMPLAINT (B)(4) HAS BEEN EVALUATED. THE BATCH 6002539 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. INVESTIGATION PROCESS OF THE COMPLAINT WAS CARRIED OUT IN ACCORDANCE WITH WORK INSTRUCTIONS GUIDELINES FOR TEST OF REFERENCE SAMPLES BUID-UMD VERSION 11 FOR THE CODE TUBING DETACHED FROM HUB. COMPLAINT INVESTIGATIONS. PHOTO/SAMPLE WAS NOT PROVIDED. IN ORDER TO TEST THE PRODUCT, THE REFERENCE SAMPLES FROM THE LOT HAS BEEN REQUESTED. THE FOLLOWING TEST WAS PERFORMED: VISUAL TEST ACCORDING TO WITH WORK INSTRUCTIONS VERSION 16 TEST ON RETURNED UNUSED/USED/REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL TEST 1 ACCORDING TO WITH WORK INSTRUCTIONS VERSION 4 TEST ON RETURNED UNUSED/USED/REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. A BATCH RECORD REVIEW WAS CONDUCTED RESULTING IN THE FOLLOWING: THE LOT 6002539 WAS MANUFACTURED ACCORDING TO THE DOCUMENT VERSION 77 ON THE PACKING PROCESS IN THE MACHINE 12, ON 06/AUG/2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORDS (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESS HAD BEEN FULFILLED AND MET THE REQUIREMENTS. CONCLUSION SUMMARY OF THE RELATED EVENT: AS A RESULT OF THE FOLLOWING: NO REPORTED HARM, NO DEFECT ON TEST, NO NON-CONFROMANCE (NC) RAISED DURING PRODUCTION. NO FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 0

E1: PATIENT CITY: (B)(6) PATIENT COUNTRY: CANADA.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA IT WAS REPORTED THAT PATIENT FACED AN INFUSION SET TUBING DETACHED EVENT ON 16-AUG-2024. THE SITE LOCATION WAS LEG. THE LOCATION OF DETACHMENT WAS AT. INFUSION SET WAS USED FOR OVERNIGHT. NO FURTHER INFORMATION WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1563947 MINMED QUICK SET UNO QUICK-SET 110/6 SC1 MCAN FPA UNOMEDICAL A/S MMT-398A600 6002539

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown