FDA Adverse Event Malfunction Summary report: N

BD SYRINGE PLASTIPAK 10ML S/SU

MDR report key: 20220919 · Received September 15, 2024

Report

Report Number
3003916417-2024-00214
Event Type
Malfunction
Date Received
September 15, 2024
Date of Event
August 26, 2024
Report Date
October 10, 2024
Manufacturer
BECTON DICKINSON IND. CIRURGICAS LTDA
Product Code
FMF
UDI-DI
07891463000378
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PHOTO RECEIVED BY OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, A HAIR WAS OBSERVED INSIDE THE BLISTER PACK, THEREFORE THE INCIDENT IS CONFIRMED. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT 4170620, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. MANUFACTURING FOR THIS PRODUCT IS PERFORMED IN A CLEAN ROOM WHICH IS KEPT UNDER A POSITIVE PRESSURE TO REDUCE THE CHANCE OF FOREIGN MATTER. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURES. ALL INDIVIDUALS ARE REQUIRED TO FOLLOW PROPER GOWNING PROCEDURES BEFORE ENTERING THE ROOM, INCLUDING HAIR PROTECTION. THE ROOT CAUSE OF THIS INCIDENT IS DUE TO PERSONNEL NOT FOLLOWING THE PROPER GOWNING PROCEDURE. MANUFACTURING PERSONNEL HAVE BEEN NOTIFIED OF THIS INCIDENT TO INCREASE AWARENESS OF THIS MATTER.

Additional Manufacturer Narrative · 0

IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD SYRINGE PLASTIPAK 10ML S/SU HAS FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER TRANSLATED FROM PORTUGUESE TO ENGLISH: HAIR FOUND INSIDE THE SEALED PACKAGING OF THE 10ML LUER-LOCK SYRINGE 990172. OCCURRENCE HAPPENED BEFORE USE. CUSTOMER INFORMS THAT HE RECEIVED A NOTIFICATION YESTERDAY (26/08/24) ABOUT A QUALITY DEVIATION IN A 990172 SYRINGE PACKAGE WITH HAIR INSIDE. QUANTITY: (B)(4) UNIT. AVAILABLE FOR COLLECTION. DEVIATION: STRAND OF HAIR INSIDE THE SYRINGE WITH SEALED PACKAGING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1101748 BD SYRINGE PLASTIPAK 10ML S/SU PISTON SYRINGE FMF BECTON DICKINSON IND. CIRURGICAS LTDA 4170620 07891463000378

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown