FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2021977 · Received March 8, 2011

Report

Report Number
3004209178-2011-80602
Event Type
Malfunction
Date Received
March 8, 2011
Date of Event
February 16, 2011
Report Date
February 16, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NO
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED BY THE CUSTOMER THAT HIS INSULIN PUMP HAD A FROZEN DISPLAY. THE CUSTOMER ALSO STATED THAT THE BUTTONS WERE NOT RESPONDING AND THE TIME ON THE CLOCK WAS NOT ADVANCING. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722WWL

Patients

Seq Age Sex Outcome Treatment
1