FDA Adverse Event
Malfunction
Summary report: N
7900
MDR report key: 2021928
·
Received February 25, 2011
Report
- Report Number
- 9680959-2011-00578
- Event Type
- Malfunction
- Date Received
- February 25, 2011
- Date of Event
- January 21, 2011
- Report Date
- February 25, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS GMBH
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE MFR'S SVC REP PERFORMED AN ON SITE INVESTIGATION. THE 12A FUSES WERE REPLACED. THE SYSTEM WAS TESTED AND IS OPERATING AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE AMPLIFIER ON THE 7900 SYSTEM DOES NOT TURN ON. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 7900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS GMBH | 7900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |