FDA Adverse Event Injury Summary report: N

INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 20219086 · Received September 14, 2024

Report

Report Number
9617229-2024-20702
Event Type
Injury
Date Received
September 14, 2024
Date of Event
July 31, 2024
Report Date
November 19, 2024
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: D9, H3, H6. LABORATORY ANALYSIS SUMMARY: THE DEVICE RELATED TO THE REPORTED EVENTS CAPSULAR CONTRACTURE WAS RECEIVED ON OCTOBER 03, 2024, WITH LOT NUMBER 1190286. BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS ARE: CAPSULAR CONTRACTURE: UNABLE TO OBSERVE. AS PER THE INVESTIGATION PROCEDURE: CREASES AND PARTICLES WERE COMPLETED. AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. THE EVENT OF "CAPSULAR CONTRACTURE" IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: CAPSULAR CONTRACTURE, BAKER GRADE III.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED, LEFT-SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE WAS EXPLANTED AND REPLACED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT-SIDE CAPSULAR CONTRACTURE BAKER GRADE III. DEVICE WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1638763 INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 1190286

Patients

Seq Age Sex Outcome Treatment
1 47 YR Female Required Intervention