FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2021891 · Received February 25, 2011

Report

Report Number
1720753-2011-01696
Event Type
Malfunction
Date Received
February 25, 2011
Date of Event
February 2, 2011
Report Date
February 25, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE GE REP PERFORMED AN ON SITE INVESTIGATION. THE FULL SET OF SOFTWARE WAS INSTALLED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED, THE SYSTEM SCREEN WAS DISPLAYING THE ERROR MESSAGE MOTION ERROR. THERE WAS NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC XRAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1