FDA Adverse Event Malfunction Summary report: N

BD SAFETYGLIDE

MDR report key: 20218730 · Received September 14, 2024

Report

Report Number
1213809-2024-00623
Event Type
Malfunction
Date Received
September 14, 2024
Date of Event
August 20, 2024
Report Date
December 4, 2024
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMI
UDI-DI
30382903059165
PMA / PMN Number
K951254
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4) - SUPPLEMENTAL MDR - NEEDLE CLOGGED/BLOCKED. NO SAMPLES OR PHOTOS RECEIVED BY OUR QUALITY TEAM FOR EVALUATION. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED BATCH NUMBER 2024137. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. PREVIOUS INVESTIGATIONS HAVE REVEALED THAT PROCESS VARIATIONS DURING THE NEEDLE LUBRICANT APPLICATION CAN CREATE CLOGGED NEEDLES. CORRECTIVE AND PREVENTATIVE ACTIONS HAVE BEEN IMPLEMENTED. SEVERAL QUALITY INITIATIVES HAVE BEEN IMPLEMENTED ON OUR MANUFACTURING LINE TO ENSURE THAT THE NEEDLE LUBRICANT APPLICATION IS PROPERLY APPLIED DURING THE MANUFACTURING PROCESS. ADDITIONALLY, A MONITORING PROGRAM IS ALSO IN PLACE TO VERIFY THE NEEDLE LUBRICANT IS APPLIED UNIFORMLY TO THE NEEDLE.

Additional Manufacturer Narrative · 0

IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION WAS PROVIDED.

Description of Event or Problem · 0

MATERIAL: 305916 BATCH#: 2024137, 3233826 IT WAS REPORTED THAT THE BD SAFETY GLIDE NEEDLE WAS CLOGGED / BLOCKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. WE ARE HEARING FROM STORES ABOUT BD SAFETYGLIDE INJECTION NEEDLES REF 305916 NOT WORKING. WHEN YOU PUSH THE PLUNGER, NOTHING COMES OUT, THERE IS TENSION. THERE SEEMS TO BE NO HOLE IN THE NEEDLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1263562 BD SAFETYGLIDE NEEDLE, HYPODERMIC, SINGLE LUMEN FMI BECTON DICKINSON MEDICAL SYSTEMS 2024137 30382903059165

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown