BD SAFETYGLIDE
Report
- Report Number
- 1213809-2024-00623
- Event Type
- Malfunction
- Date Received
- September 14, 2024
- Date of Event
- August 20, 2024
- Report Date
- December 4, 2024
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- FMI
- UDI-DI
- 30382903059165
- PMA / PMN Number
- K951254
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
(B)(4) - SUPPLEMENTAL MDR - NEEDLE CLOGGED/BLOCKED. NO SAMPLES OR PHOTOS RECEIVED BY OUR QUALITY TEAM FOR EVALUATION. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED BATCH NUMBER 2024137. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. PREVIOUS INVESTIGATIONS HAVE REVEALED THAT PROCESS VARIATIONS DURING THE NEEDLE LUBRICANT APPLICATION CAN CREATE CLOGGED NEEDLES. CORRECTIVE AND PREVENTATIVE ACTIONS HAVE BEEN IMPLEMENTED. SEVERAL QUALITY INITIATIVES HAVE BEEN IMPLEMENTED ON OUR MANUFACTURING LINE TO ENSURE THAT THE NEEDLE LUBRICANT APPLICATION IS PROPERLY APPLIED DURING THE MANUFACTURING PROCESS. ADDITIONALLY, A MONITORING PROGRAM IS ALSO IN PLACE TO VERIFY THE NEEDLE LUBRICANT IS APPLIED UNIFORMLY TO THE NEEDLE.
IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.
NO ADDITIONAL INFORMATION WAS PROVIDED.
MATERIAL: 305916 BATCH#: 2024137, 3233826 IT WAS REPORTED THAT THE BD SAFETY GLIDE NEEDLE WAS CLOGGED / BLOCKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. WE ARE HEARING FROM STORES ABOUT BD SAFETYGLIDE INJECTION NEEDLES REF 305916 NOT WORKING. WHEN YOU PUSH THE PLUNGER, NOTHING COMES OUT, THERE IS TENSION. THERE SEEMS TO BE NO HOLE IN THE NEEDLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1263562 | BD SAFETYGLIDE | NEEDLE, HYPODERMIC, SINGLE LUMEN | FMI | BECTON DICKINSON MEDICAL SYSTEMS | 2024137 | 30382903059165 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |