FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2021859 · Received February 25, 2011

Report

Report Number
1720753-2011-01674
Event Type
Malfunction
Date Received
February 25, 2011
Date of Event
January 27, 2011
Report Date
February 25, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE REP RESEATED A CIRCUIT BOARD AND CONNECTORS. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE 9800 SYSTEM DISPLAYED CINE ERROR FAILURE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPY X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1