FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2021657 · Received February 21, 2011

Report

Report Number
1720753-2011-01537
Event Type
Malfunction
Date Received
February 21, 2011
Date of Event
February 1, 2011
Report Date
February 21, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE RTOS SINGLEBOARD COMPUTER. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED, THE SYSTEM DISPLAYED A "INVALID INPUT SIGNAL" ERROR MESSAGE AND WOULD NOT COMPLETE BOOT UP. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1