FDA Adverse Event Malfunction Summary report: N

6800

MDR report key: 2021611 · Received February 21, 2011

Report

Report Number
1720753-2011-01510
Event Type
Malfunction
Date Received
February 21, 2011
Date of Event
February 9, 2011
Report Date
February 21, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE FOOT SWITCH MAY HAVE BEEN PRESSED DURING BOOT UP CAUSING THE ERROR. THE FOOT SWITCH WAS REPOSITIONED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 6800 SYSTEM HAD AN X-RAY SWITCH ERROR AT BOOT UP. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 6800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 6800

Patients

Seq Age Sex Outcome Treatment
1