FDA Adverse Event Malfunction Summary report: N

REFURBED COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNE

MDR report key: 2021484 · Received March 17, 2011

Report

Report Number
6000001-2011-01923
Event Type
Malfunction
Date Received
March 17, 2011
Date of Event
March 1, 2011
Report Date
March 1, 2011
Manufacturer
BAXTER HEALTHCARE - BUFFALO GROVE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 550:320:844:0000 THAT FAILED TO AUDIBLY ALARM WAS CONFIRMED BY BAXTER PERSONNEL DURING PRODUCT EVALUATION. THE ROOT CAUSE OF THIS CONDITION WAS NOT IDENTIFIED. THE HARNESS VOLUME CONTROL WAS REPLACED AS A PRECAUTION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM.? THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH MDQ-CAPA (B)(4). A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION.

Description of Event or Problem · 1

A BAXTER SERVICE TECHNICIAN DISCOVERED A COLLEAGUE INFUSION PUMP EXPERIENCED FAILURE CODE 550:320:844:0000. THIS PROBLEM WAS IDENTIFIED DURING SERVICE AND FAILED TO AUDIBLY ALARM AS THE PUMP WAS BEING TESTED. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.09.90.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 REFURBED COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNE PUMP, INFUSION FRN BAXTER HEALTHCARE - BUFFALO GROVE

Patients

Seq Age Sex Outcome Treatment
1