FDA Adverse Event Malfunction Summary report: N

BD PRESET¿

MDR report key: 20214770 · Received September 13, 2024

Report

Report Number
9617032-2024-01462
Event Type
Malfunction
Date Received
September 13, 2024
Date of Event
August 16, 2024
Report Date
August 21, 2024
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
UDI-DI
50382903644150
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: "MATERIAL #: 364415 LOT/BATCH #: 3251818 . BD HAD NOT RECEIVED SAMPLES, BUT 3 PHOTOS WERE PROVIDED FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND THE INDICATED FAILURE MODE FOR FOREIGN MATTER WAS OBSERVED. THE CANNULA HAS EPOXY DRIP OVER THE NEEDLE TIP. ADDITIONALLY, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND THE ISSUE OF FOREIGN MATTER WAS NOT OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE FOREIGN MATTER. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS."

Description of Event or Problem · 0

IT WAS REPORTED PRIOR TO USING A BD PRESET¿ SYRINGE, FOREIGN MATTER WAS FOUND ON THE CANNULA. THE DEVICE WAS NOT USED. THERE WAS NO REPORT OF ADVERSE IMPACT TO PATIENTS OR USERS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1536062 BD PRESET¿ BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 3251818 50382903644150

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown