FDA Adverse Event Malfunction Summary report: N

CAMERA HEAD AC - C-MOUNT

MDR report key: 20213984 · Received September 13, 2024

Report

Report Number
1221934-2024-03190
Event Type
Malfunction
Date Received
September 13, 2024
Date of Event
August 30, 2024
Report Date
September 13, 2024
Manufacturer
MEDOS INTERNATIONAL SARL
Product Code
FWF
UDI-DI
10886705028733
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. H4: DEVICE MANUFACTURE DATE HAS BEEN ADDED.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. H4: THE DEVICE MANUFACTURE DATE WAS UNKNOWN.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. H11 ADDITIONAL NARRATIVE: THE COMPLAINT DEVICE WAS RECEIVED AT THE SERVICE CENTER AND EVALUATED. DURING THE SERVICE EVALUATION THE FOLLOWING DEFECTS WERE IDENTIFIED: FUNCTIONAL : INTERMITTENT OPERATION PER SERVICE REPORTS, THIS COMPLAINT WAS CONFIRMED. BASED ON SERVICE MANUAL OPERATIONAL AND DIAGNOSTIC, THE DEVICE WAS FOUND TO BE NON-REPAIRABLE, AND IT WAS PLACED INTO LONG TERM HOLD. BASED ON THE INVESTIGATION FINDINGS, THE FAULTY PARTS WERE IDENTIFIED AS THE ROOT CAUSE FOR THE DEVICE FAILURE DURING THE SERVICE EVALUATION. IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED AND THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF J&J MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

IT WAS REPORTED BY THE SALES REP THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE ON (B)(6) 2024 THE CAMERA HEAD AC - C-MOUNT DEVICE LIGHT WAS FLUCTUATING THROUGHOUT. THE DEVICE ALSO CUT OUT TWICE DURING THE CASE BUT CAME BACK ON TWO TO THREE SECONDS AFTER GOING OUT. THE CASE WAS COMPLETED WITH THE SAME CAMERA WITHOUT ISSUE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1563441 CAMERA HEAD AC - C-MOUNT CAMERA, TELEVISION, ENDOSCOPIC, WITHOUT AUDIO FWF MEDOS INTERNATIONAL SARL 10886705028733

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown