PENUMBRA SYSTEM ASPIRATION PUMP
Report
- Report Number
- 3005168196-2011-00026
- Event Type
- Malfunction
- Date Received
- February 8, 2011
- Date of Event
- January 11, 2011
- Report Date
- January 12, 2011
- Manufacturer
- PENUMBRA, INC.
- Product Code
- BTA
- PMA / PMN Number
- K051758
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
DEVICE INVESTIGATION: THE PUMP HAS A BENT LOWER CHASSIS AT ONE OF THE MOUNTING SCREWS. THE REST OF THE EXTERIOR APPEARS NORMAL FOR A USED PUMP. OPENING THE CHASSIS THERE IS NOTHING VISIBLY UNUSUAL INSIDE THE PUMP. RESULTS: THE PUMP WAS PLUGGED IN AND TURNED ON. THE PUMP FUNCTIONED AS EXPECTED. THE PRESSURE ADJUSTMENT KNOB WAS TURNED AND MOVED FREELY. THE PUMP WAS RETURNED TO PENUMBRA ADJUSTED FOR MAXIMUM VACUUM, WHICH MEASURED >30 IN HG ON THE PUMP GAUGE. BASED ON THE DESCRIPTION OF THE COMPLAINT (BURNING RUBBER AND VISIBLE SMOKE), THE UNIT WAS RUN FOR 15 MINUTES UNDER CLOSE OBSERVATION. NO SMOKE WAS OBSERVED NOR WAS ANY BURNT RUBBER SMELL PERCEIVED. TO FURTHER STRESS THE UNIT AND SIMULATE WORST CASE CONDITIONS, THE UNIT WAS PLACED ON A FOLDED SURGICAL DRAPE TO BLOCK THE INLET AIR VENT AND RUN AT MAXIMUM VACUUM FOR 15 MINUTES. AGAIN NO SMOKE WAS OBSERVED AND NO BURNT RUBBER SMELL WAS PERCEIVED. THE METAL ON THE PUMP MOTOR HOUSING WAS WARM TO THE TOUCH BUT NOT UNEXPECTED. CONCLUSION CODE: WE WERE UNABLE TO CONFIRM THE COMPLAINT. BASED ON THE DAMAGE TO THE LOWER CHASSIS OF THE PUMP, IT APPEARS THAT THE PUMP MAY HAVE BEEN DAMAGED AT SOME POINT. THIS DAMAGE DOES NOT AFFECT THE PERFORMANCE OF THE PUMP.
THE LAB MGR REPORTED THAT THE PENUMBRA SYSTEM ASPIRATION PUMP BEGAN TO SMELL LIKE BURNING RUBBER DURING A STROKE CASE, AND THE NURSING NOTED SMOKE COMING FROM THE PUMP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PENUMBRA SYSTEM ASPIRATION PUMP | BTA & JCX | BTA | PENUMBRA, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |