PENUMBRA SYSTEM ASPIRATION PUMP
Report
- Report Number
- 3005168196-2011-00023
- Event Type
- Malfunction
- Date Received
- February 8, 2011
- Date of Event
- January 7, 2011
- Report Date
- January 11, 2011
- Manufacturer
- PENUMBRA, INC.
- Product Code
- BTA
- PMA / PMN Number
- K051758
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
DEVICE INVESTIGATION: THE PUMP WAS EXTERNALLY LABELED AS NON-FUNCTIONAL BY THE HOSPITAL. THERE WAS NOTHING UNUSUAL ABOUT THE APPEARANCE OF THE PUMP. EVAL, RESULTS: THE PUMP WAS PLUGGED IN AND TURNED ON. THE PRESSURE ADJUSTMENT VALVE WAS RETURNED TO PENUMBRA ADJUSTED SIGNIFICANTLY BELOW MAXIMUM VACUUM. THE VACUUM IN ITS RETURNED STATED WAS -5 IN HG. THE PRESSURE KNOB WAS TWO TURNS COUNTERCLOCKWISE FROM THE MAXIMUM VACUUM ADJUSTMENT. ONCE THE PRESSURE VALVE WAS ADJUSTED, THE VACUUM MEASURED -27.5 IN HG. THE PUMP IS FULLY FUNCTIONAL. THE PUMP VACUUM AS DESCRIBED IN THE COMPLAINT WAS INITIALLY CONFIRMED. HOWEVER, ONCE THE PUMP PRESSURE KNOB WAS CORRECTLY ADJUSTED IT WORKED WITHIN SPECIFICATION.
DURING A STROKE CASE THE PENUMBRA SYSTEM ASPIRATION PUMP DID NOT HAVE ANY SUCTION. A SECOND PUMP WAS USED TO COMPLETE THE CASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PENUMBRA SYSTEM ASPIRATION PUMP | BTA & JCX | BTA | PENUMBRA, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |