FDA Adverse Event Malfunction Summary report: N

VIDEO-OPTIK "ENDOEYE 3D", 30°

MDR report key: 20212878 · Received September 13, 2024

Report

Report Number
9610773-2024-32548
Event Type
Malfunction
Date Received
September 13, 2024
Report Date
May 12, 2025
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
GCM
PMA / PMN Number
K193026
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SENT FOR A CORRECTION TO THE INITIAL REPORT.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE FINAL INVESTIGATION. UPDATED FIELDS: D8, H2, H3, H6, H11 THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND THE REPORTED FAILURE WAS CONFIRMED. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: COMPONENT FAILURE SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Additional Manufacturer Narrative · 0

E2/E3: NO INFORMATION AVAILABLE FOR THE OCCUPATION OF THE INITIAL REPORTER OR WHETHER THEY ARE A HEALTHCARE PROFESSIONAL. THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE RIGID VIDEO SCOPE VIDEO IMAGE BECAME WHITE. THERE WERE NO REPORTS OF PATIENT HARM.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE RIGID VIDEO SCOPE VIDEO IMAGE BECAME WHITE. THERE WERE NO REPORTS OF PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1101244 VIDEO-OPTIK "ENDOEYE 3D", 30° RIGID VIDEO LAPAROSCOPE GCM OLYMPUS WINTER & IBE GMBH WA50082A

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown