SYNCHRON® LX RHEUMATOID FACTOR (RF REAGENT)
Report
- Report Number
- 2050012-2011-00728
- Event Type
- Malfunction
- Date Received
- March 17, 2011
- Date of Event
- January 20, 2011
- Report Date
- February 14, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- DHR
- PMA / PMN Number
- K971788
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
CUSTOMER STATED THAT QC IS RECOVERING WITHIN THE ACCEPTABLE RANGE. NO SYSTEM ISSUES WERE NOTED BY THE CUSTOMER. THIS EVENT IS A REAGENT RELATED ISSUE. THUS, SERVICE WAS NOT NEEDED FOR THIS EVENT. A NEWER LOT OF RF REAGENT WILL BE SENT TO THE CUSTOMER.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO OBTAINING A FALSELY POSITIVE RHEUMATOID FACTOR (RF) RESULT GENERATED BY THE UNICEL DXC 600 PRO SYNCHRON CHEMISTRY ANALYZER. THE ERRONEOUS RESULT WAS REPORTED OUT OF THE LABORATORY. THE SAMPLE WAS RETESTED WITH AN ALTERNATE CALIBRATOR THAT CONFIRMED THE SAMPLE TO YIELD A NEGATIVE RF RESULT. THE PATIENT WAS REFERRED TO A RHEUMATOLOGIST AS A RESULT OF THIS EVENT. THIS REPORT IS RELATED TO THE FOLLOWING MDRS THAT ARE BEING REPORTED ON DIFFERENT DATES FOR THE SIMILAR EVENTS THAT OCCURRED AT THIS CUSTOMER SITE: 2050012-2011-00725, 2050012-2011-00726, 2050012-2011-00763, 2050012-2011-00727, 2050012-2011-00729, 2050012-2011-00730, 2050012-2011-00731, 2050012-2011-00733, 2050012-2011-00734, 2050012-2011-00735, 2050012-2011-00737, 2050012-2011-00738.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON® LX RHEUMATOID FACTOR (RF REAGENT) | RHEUMATOID FACTOR REAGENT | DHR | BECKMAN COULTER INC. | NA | M004772 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |