FDA Adverse Event Malfunction Summary report: N

LENSX LASER SYSTEM

MDR report key: 20210524 · Received September 13, 2024

Report

Report Number
2028159-2024-01364
Event Type
Malfunction
Date Received
September 13, 2024
Report Date
November 18, 2024
Manufacturer
ALCON LENSX, INC.
Product Code
OOE
UDI-DI
00380659981623
PMA / PMN Number
K163551
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

A MANUFACTURING DEVICE HISTORY RECORD REVIEW WAS PERFORMED PRIOR TO PRODUCT RELEASE TO ENSURE THAT THE PRODUCT WAS MANUFACTURED IN COMPLIANCE WITH THE DEVICE MASTER RECORD. BASED ON THE ASSESSMENT, THE PRODUCT MET RELEASE CRITERIA. THE COMPANY REPRESENTATIVE WAS UNABLE TO CONFIRM NOR REPLICATE THE REPORTED EVENT. THE SYSTEM WAS TESTED AND FOUND TO MEET PRODUCT SPECIFICATIONS. THE CUSTOMER REPORTED EVENT CANNOT BE CONFIRMED. THUS, BASED ON THE INFORMATION OBTAINED, THE ROOT CAUSE OF THE REPORTED EVENT IS INCONCLUSIVE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 0

H.6.: INVESTIGATION, INCLUDING ROOT CAUSE ANALYSIS, IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. ALCON LENSX (SITE #3008772169) IS NO LONGER OPERATIONAL. LENSX MANUFACTURED PRODUCTS ARE MAINTAINED AND INVESTIGATED BY THE ALCON RESEARCH, LTD. IRVINE TECHNOLOGY CENTER SITE #2028159). THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 0

A NON-HEALTHCARE PROFESSIONAL REPORTED THAT FLAP ISSUE IN THE UNKNOWN EYE, DURING FLAP CREATION PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1564234 LENSX LASER SYSTEM OPHTHALMIC FEMTOSECOND LASER OOE ALCON LENSX, INC. 550 NA 00380659981623

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown