FDA Adverse Event Summary report: N

UNICEL® DXC 800 PRO SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 2021008 · Received March 16, 2011

Report

Report Number
2050012-2011-00753
Date Received
March 16, 2011
Date of Event
February 15, 2011
Report Date
February 15, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A BCI CUSTOMER TECHNICAL SUPPORT(CTS) ASSISTED THE CUSTOMER HAD CUSTOMER CHECK VACUUM ACCUMULATOR FOR FLUID. CTS ALSO INSTRUCTED THE CUSTOMER TO TIGHTEN THE VACUUM ACCUMULATOR CANISTER. A BCI FIELD SERVICE ENGINEER (FSE) REPLACED BOTH VACUUM AND PRESSURE PUMPS.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT A NO FOAM BOTTLE LEAK. NO INJURY WAS REPORTED OR OCCURRED DUE TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 800 PRO SYNCHRON ® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 800 PRO N/A

Patients

Seq Age Sex Outcome Treatment
1