UNICEL® DXC 800 PRO SYNCHRON ® CHEMISTRY ANALYZER
MDR report key: 2020880
·
Received March 16, 2011
Report
- Report Number
- 2050012-2011-00751
- Date Received
- March 16, 2011
- Date of Event
- February 14, 2011
- Report Date
- February 14, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE CUSTOMER CHANGED THE CL ELECTRODE TIP WITHOUT ADHERING TO THE PROPER MAINTENANCE PROCEDURE AND WITHOUT DRAINING THE FLOWCELL. THIS CAUSED A FEW DROPS OF REAGENTS LEAK FROM THE FLOWCELL, BUT CUSTOMER WAS WEARING PPE. SERVICE WAS NOT DISPATCHED.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT ION-SELECTIVE ELECTRODE (ISE) MODULE LEAK. NO INJURY WAS REPORTED.