UNICEL® DXC 800 PRO SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 2020880 · Received March 16, 2011

Report

Report Number
2050012-2011-00751
Date Received
March 16, 2011
Date of Event
February 14, 2011
Report Date
February 14, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER CHANGED THE CL ELECTRODE TIP WITHOUT ADHERING TO THE PROPER MAINTENANCE PROCEDURE AND WITHOUT DRAINING THE FLOWCELL. THIS CAUSED A FEW DROPS OF REAGENTS LEAK FROM THE FLOWCELL, BUT CUSTOMER WAS WEARING PPE. SERVICE WAS NOT DISPATCHED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT ION-SELECTIVE ELECTRODE (ISE) MODULE LEAK. NO INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 800 PRO SYNCHRON ® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 800 PRO N/A

Patients

Seq Age Sex Outcome Treatment
1