FDA Adverse Event Injury Summary report: N

MULTI-LINK OTW VISION CORONARY STENT SYSTEM

MDR report key: 2020743 · Received March 16, 2011

Report

Report Number
2024168-2011-01769
Event Type
Injury
Date Received
March 16, 2011
Date of Event
February 18, 2011
Report Date
February 21, 2011
Manufacturer
AV-TEMECULA-CT
Product Code
MAF
PMA / PMN Number
P020047
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE MULTI LINK RX VISION STENT IS BEING FILED UNDER A SEPARATE MEDWATCH MFR NUMBER. THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. THE INABILITY TO CROSS A LESION CAN BE AFFECTED BY NUMEROUS FACTORS INCLUDING, BUT ARE NOT LIMITED TO, PATIENT ANATOMICAL MORPHOLOGY, PATIENT DISEASE STATE, PRE-DILATATION STRATEGY, DEVICE PLACEMENT TECHNIQUE, AND ACCESSORY DEVICE SUPPORT. ALTHOUGH THE ANATOMICAL CONDITIONS WERE NOT REPORTED, FAILURE TO ADVANCE IS NOT OFTEN ASSOCIATED WITH A PRODUCT QUALITY DEFICIENCY AND IS LIKELY RELATED TO THE OPERATIONAL CONTEXT OF THE PROCEDURE. THE REPORTED DISSECTION IS A KNOWN ADVERSE EVENT LISTED IN THE VISION INSTRUCTIONS FOR USE. ALTHOUGH A CONCLUSIVE CAUSE FOR THE REPORTED PATIENT EFFECT AND THE RELATIONSHIP TO THE DEVICE, IF ANY, CANNOT BE DETERMINED, THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING. THE REPORTED FAILURE TO ADVANCE APPEARS TO BE RELATED TO THE OPERATIONAL CONTEXT OF THE PROCEDURE. A REVIEW OF THE FINISHED DEVICE LOT HISTORY RECORDS DID NOT REVEAL ANY NONCONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT WHICH COULD HAVE CONTRIBUTED TO THE REPORTED EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER TWO VISION STENT DELIVERY SYSTEMS (2.75X08 MM RX VISION, 2.75X12 OTW VISION) FAILED TO CROSS TO TREAT A LESION LOCATED IN THE CIRCUMFLEX, A DISSECTION WAS NOTED; HOWEVER, IT WAS UNKNOWN WHICH DEVICE CAUSED THE DISSECTION. THE DISSECTION WAS SUCCESSFULLY TREATED WITH A BALLOON DILATATION CATHETER AND THE PATIENT WAS FINE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MULTI-LINK OTW VISION CORONARY STENT SYSTEM CORONARY STENT SYSTEM MAF AV-TEMECULA-CT 9120241

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention