INTROCAN SAFETY ® 2
Report
- Report Number
- 9610825-2024-00721
- Event Type
- Malfunction
- Date Received
- September 12, 2024
- Report Date
- October 10, 2024
- Manufacturer
- B. BRAUN MELSUNGEN AG
- Product Code
- FOZ
- PMA / PMN Number
- K220756
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INTERNAL REPORT NUMBER (B)(4). EVENT 2: WHILE A SAMPLE OF THE INTROCAN SAFETY 2 WL PUR M 24GX19MM - US WAS NOT AVAILABLE, A BD Q-STYLE EXTENSION SET WAS SUBMITTED TO THE MANUFACTURER FOR EVALUATION. IT IS NOTED THAT THE BD Q-STYLE EXTENSION SET IS NOT A B. BRAUN MEDICAL INC. PRODUCT. THROUGH MICROSCOPIC EXAMINATION, NO DAMAGES WERE OBSERVED ON THE LUER LOCK CONNECTOR OF THE EXTENSION SET. A SIMULATION WAS PERFORMED ON A FRESH INTROCAN SAFETY 2-M G24 CATHETER HUB SAMPLE AND WAS CONNECTED TO THE RETURNED BD Q-STYLE EXTENSION SET AND FLUSHED WITH RED DYE WATER TO CHECK FOR LEAKAGE AT THE LUER CONNECTION AREA. NO LEAKAGE WAS OBSERVED FROM THE LUER CONNECTION BETWEEN EXTENSION SET AND THE CATHETER HUB. FURTHER TESTS WAS ALSO CONDUCTED ON A FRESH SAMPLE OF AN INTROCAN AND INTROCAN SAFETY 3; NO LEAKAGE WAS OBSERVED. BASED ON THE SIMULATION, IT WAS CONFIRMED THAT THE B. BRAUN IV CATHETERS ARE COMPATIBLE WITH THE RETURNED BD Q-STYLE EXTENSION SET AS NO LEAKAGE WAS OBSERVED BETWEEN THE CONNECTORS. BASED ON THE INVESTIGATION, THE REPORTED DEFECT WAS NOT ABLE TO BE OBSERVED OR REPLICATED. WE WILL MAINTAIN THIS REPORT FOR FURTHER REFERENCES AND CONTINUE TO MONITOR OTHER REPORTS FOR SIMILAR OCCURRENCES. IF ANY ADDITIONAL PERTINENT INFORMATION BECOMES AVAILABLE, A FOLLOW UP WILL BE SUBMITTED.
THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INTERNAL REPORT NUMBER (B)(4). THE INVESTIGATION IS ONGOING AT THIS TIME. A FOLLOW-UP WILL BE SUBMITTED WHEN THE INVESTIGATION RESULTS BECOME AVAILABLE.
AS REPORTED BY THE USER FACILITY: EVENT 2. BRIEF INQUIRY DESCRIPTION: CATHETER LEAKING AT CONNECTION TO EXTENSION SET. DETAILED INQUIRY DESCRIPTION: IV CATHETER IS LEAKING AT CONNECTION SITE TO EXTENSION SET. EXTENSION SET IS BD Q-SYTE, NOT A BBRAUN PRODUCT. CUSTOMER REPORTED THAT THE IV HUB IS NOT CONNECTING WELL OR IT IS WORKING ITS WAY LOOSE, ALLOWING FLUID TO LEAK OUT UNDER THE IV DRESSING DURING THERAPY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1516722 | INTROCAN SAFETY ® 2 | CATHETER,INTRAVASCULAR,THE | FOZ | B. BRAUN MELSUNGEN AG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |