FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM

MDR report key: 2020687 · Received March 16, 2011

Report

Report Number
3005075853-2011-01045
Event Type
Malfunction
Date Received
March 16, 2011
Date of Event
February 19, 2011
Report Date
February 21, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). EMPTY. THE ANALYSIS RESULTS FOUND THAT DEVICE (A) WAS RECEIVED IN GOOD VISUAL CONDITION. UPON CYCLING THE DEVICE, IT WAS NOTED TO BE EMPTY AND THE LOCKOUT HAD BEEN FIRED THROUGH. PLEASE NOTE THE DEVICE CONTAINS A LOCKOUT FEATURE THAT IS DESIGNED TO INCREASE THE REQUIRED FORCE IT TAKES TO CLOSE THE TRIGGER ONCE THE LAST CLIP HAS BEEN FIRED; THIS REDUCES THE POSSIBILITY OF EMPTY JAWS BEING CLOSED ON A STRUCTURE. IN ADDITION, IF THE TRIGGER IS FORCED CLOSED ONCE THE LOCKOUT HAS BEEN ENGAGED, THE JAWS MAY REMAIN IN THE CLOSED POSITION.THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS. THE ANALYSIS RESULTS FOUND THAT DEVICE (B) WAS RECEIVED IN GOOD VISUAL CONDITION. UPON CYCLING THE DEVICE, IT WAS NOTED TO BE EMPTY AND THE LOCKOUT HAD BEEN FIRED THROUGH. PLEASE NOTE THE DEVICE CONTAINS A LOCKOUT FEATURE THAT IS DESIGNED TO INCREASE THE REQUIRED FORCE IT TAKES TO CLOSE THE TRIGGER ONCE THE LAST CLIP HAS BEEN FIRED; THIS REDUCES THE POSSIBILITY OF EMPTY JAWS BEING CLOSED ON A STRUCTURE. IN ADDITION, IF THE TRIGGER IS FORCED CLOSED ONCE THE LOCKOUT HAS BEEN ENGAGED, THE JAWS MAY REMAIN IN THE CLOSED POSITION. NO INCIDENT RELATED TO THE REPORTED EVENT WAS OBSERVED DURING THE BATCH RECORD REVIEW. BATCH # G9L82R MFG DATE 10/12/2010; EXP DATE 9/12/2015.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY WHILE LIGATING THE CYSTIC ARTERY, THE JAWS DETACHED. THERE WAS NO PATIENT CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-5MM CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA G4TP3F

Patients

Seq Age Sex Outcome Treatment
1