FDA Adverse Event Malfunction Summary report: N

ARCHITECT I2000SR ANALYZER

MDR report key: 2020631 · Received March 16, 2011

Report

Report Number
1628664-2011-00112
Event Type
Malfunction
Date Received
March 16, 2011
Date of Event
February 23, 2011
Report Date
February 24, 2011
Manufacturer
ABBOTT MANUFACTURING, INC.
Product Code
JJE
PMA / PMN Number
K983212
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN EVALUATION WAS PERFORMED WHICH INCLUDED REVIEW OF COMPLAINT INFORMATION, INSTRUMENT HISTORY, FIELD COMPLAINT REVIEW, AND REVIEW OF ASSOCIATED LABELING. THE CUSTOMER REPLACED THE ARCHITECT I2000SR PROBE SINCE IT WAS APPROXIMATELY 4 MONTHS OLD. THE OPERATOR VALIDATED THE BHCG ASSAY BY PERFORMING REPRODUCIBILITY TESTS. HE PERFORMED THE APPROPRIATE TROUBLESHOOTING STEPS AND THE COMPLAINT ISSUE WAS RESOLVED. QUALITY METRICS WERE REVIEWED AND NO ADVERSE TRENDS WERE IDENTIFIED RELATED TO THE CUSTOMER ISSUE. BASED UPON THE AVAILABLE INFORMATION ALONG WITH THE RESULTS OF THIS PRODUCT EVALUATION OF THE ARCHITECT I2000SR SYSTEM, NO PRODUCT DEFICIENCY OF THE ARCHITECT I2000SR INSTRUMENT WAS FOUND.

Additional Manufacturer Narrative · 1

(B)(4). AN INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT A POSITIVE ARCHITECT BHCG RESULT OF 30 MIU/ML WAS GENERATED FOR A PATIENT SAMPLE THAT RETESTED NEGATIVE AT <1.2 MIU/ML THREE TIMES. CONTROLS WERE WITHIN RANGE. THE PATIENT SAMPLE TESTED PRIOR TO THIS SAMPLE HAD A HIGH BHCG CONCENTRATION OF 100,000 MIU/ML. THE CUSTOMER CHANGED THE SAMPLE PROBE ON THE ARCHITECT (B)(4) TO RESOLVE THE ISSUE. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARCHITECT I2000SR ANALYZER AUTOMATED IMMUNOASSAY ANALYZER JJE ABBOTT MANUFACTURING, INC.

Patients

Seq Age Sex Outcome Treatment
1 ARCHITECT BHCG REAGENT LIST 7K78-25, LOT 94908JN00| ARCHITECT BHCG REAGENT LIST 7K78-25, LOT 94908JN00