ARCHITECT I2000SR ANALYZER
Report
- Report Number
- 1628664-2011-00112
- Event Type
- Malfunction
- Date Received
- March 16, 2011
- Date of Event
- February 23, 2011
- Report Date
- February 24, 2011
- Manufacturer
- ABBOTT MANUFACTURING, INC.
- Product Code
- JJE
- PMA / PMN Number
- K983212
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). AN EVALUATION WAS PERFORMED WHICH INCLUDED REVIEW OF COMPLAINT INFORMATION, INSTRUMENT HISTORY, FIELD COMPLAINT REVIEW, AND REVIEW OF ASSOCIATED LABELING. THE CUSTOMER REPLACED THE ARCHITECT I2000SR PROBE SINCE IT WAS APPROXIMATELY 4 MONTHS OLD. THE OPERATOR VALIDATED THE BHCG ASSAY BY PERFORMING REPRODUCIBILITY TESTS. HE PERFORMED THE APPROPRIATE TROUBLESHOOTING STEPS AND THE COMPLAINT ISSUE WAS RESOLVED. QUALITY METRICS WERE REVIEWED AND NO ADVERSE TRENDS WERE IDENTIFIED RELATED TO THE CUSTOMER ISSUE. BASED UPON THE AVAILABLE INFORMATION ALONG WITH THE RESULTS OF THIS PRODUCT EVALUATION OF THE ARCHITECT I2000SR SYSTEM, NO PRODUCT DEFICIENCY OF THE ARCHITECT I2000SR INSTRUMENT WAS FOUND.
(B)(4). AN INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
THE CUSTOMER STATED THAT A POSITIVE ARCHITECT BHCG RESULT OF 30 MIU/ML WAS GENERATED FOR A PATIENT SAMPLE THAT RETESTED NEGATIVE AT <1.2 MIU/ML THREE TIMES. CONTROLS WERE WITHIN RANGE. THE PATIENT SAMPLE TESTED PRIOR TO THIS SAMPLE HAD A HIGH BHCG CONCENTRATION OF 100,000 MIU/ML. THE CUSTOMER CHANGED THE SAMPLE PROBE ON THE ARCHITECT (B)(4) TO RESOLVE THE ISSUE. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ARCHITECT I2000SR ANALYZER | AUTOMATED IMMUNOASSAY ANALYZER | JJE | ABBOTT MANUFACTURING, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ARCHITECT BHCG REAGENT LIST 7K78-25, LOT 94908JN00| ARCHITECT BHCG REAGENT LIST 7K78-25, LOT 94908JN00 |