ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
Report
- Report Number
- 2531779-2011-00507
- Event Type
- Malfunction
- Date Received
- February 16, 2011
- Report Date
- December 30, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- UNKNOWN
Narratives
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED INTACT WITH NO LIFTING OR PEELING OBSERVED, THE UP/DOWN/OK/CONTRAST KEYPAD BUTTONS DID NOT RESPOND TO USER INPUTS DURING TESTING, IT WAS OBSERVED THAT THE CONTRAST KEY CONTACT WAS MISALIGNED, THE CONTRAST/OK KEY CONTACTS WERE INVERTED, THE UP/DOWN KEY CONTACTS BECAME INVERTED WHEN PRESSED AND DID NOT ALWAYS SPRING BACK, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION FOUND UNDER ALL KEY CONTACTS.
IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE OK KEYPAD BUTTON REPORTEDLY HAS TO BE PRESSED MULTIPLE TIMES FOR A RESPONSE. THE KEYPAD IS INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING |