FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2020560 · Received February 16, 2011

Report

Report Number
2531779-2011-00503
Event Type
Malfunction
Date Received
February 16, 2011
Report Date
December 29, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. EVALUATION OF THE PRODUCT REVEALED THAT THE PRIME HISTORY AND SUSPEND HISTORY WAS NOT SHOWING DATA CORRECTLY. A 10 UNIT "NORMAL BOLUS" WAS PERFORMED DURING INVESTIGATION AND WAS CORRECTLY CAPTURED IN THE PUMP HISTORY. A "SUSPEND" AND A "PRIME" WERE ALSO PERFORMED DURING INVESTIGATION. HOWEVER, THE PUMP HISTORY WAS OBSERVED TO BE OVERWRITTEN SO THAT ONLY THE LATEST SUSPEND AND PRIME WERE PRESENT. THE PUMP BLACK BOX DOWNLOAD DID NOT SHOW ANY ERRORS OR ALARMS. AN EEPROM CORRUPTION WAS OBSERVED.

Description of Event or Problem · 1

ON (B)(6), 2010, THE REPORTER CONTACTED ANIMAS ON BEHALF OF HER SON (THE PATIENT) ALLEGING THAT THE PUMP'S PRIME HISTORY WAS ONLY HOLDING ONE RECORD. THE REPORTER DID NOT ALLEGE ANY HARM OR INJURY BECAUSE OF THE ALLEGED ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1