FDA Adverse Event Injury Summary report: N

KINOS AXIOM TOTAL ANKLE SYSTEM

MDR report key: 20205513 · Received September 12, 2024

Report

Report Number
3014833750-2024-00002
Event Type
Injury
Date Received
September 12, 2024
Report Date
September 12, 2024
Manufacturer
RESTOR3D, INC.
Product Code
HSN
PMA / PMN Number
K192778
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

HEALTH EFFECT CODE 2646 CHOSEN BECAUSE SYSTEM DID NO RECOGNIZE INTENDED CODE 4883 AS PART OF THE INVESTIGATION, IMAGES OF THE PATIENT'S ANKLE AT THE TIME OF THE ORIGINAL SURGERY AND NOW WERE REVIEWED FOR POTENTIAL CAUSE FOR THE REPORTED TIBIAL IMPLANT LOOSENING. IN THE IMAGES, IT APPEARS THAT BONE CEMENT WAS NOT USED ON THE BONE CONTACTING SURFACE OF THE IMPLANT, ALTHOUGH THE DEVICE IS ONLY INDICATED FOR USE WITH CEMENT, PER LBL-50013 IFU -KINOS TAR. BONE CEMENT IS USED TO SECURELY AFFIX THE IMPLANT TO THE BONE; WITHOUT IT, THERE IS POTENTIAL RISK OF IMPLANT LOOSENING, AS SEEN REPORTED IN THE COMPLAINT. BASED ON THE INVESTIGATION, THE ARE NO DEFICIENCIES WITH THE DESIGN OR RISK CONSIDERATIONS OF THE IMPLANT OR THE ABILITY OF THE PLASMA SPRAY/BONE CEMENT COMBINATION TO PROVIDE SECURE FIXATION TO THE DISTAL TIBIA.

Description of Event or Problem · 0

TIBIAL IMPLANT LOOSENING ON A TOTAL ANKLE REPLACEMENT THAT WAS IMPLANTED MULTIPLE YEARS AGO REQUIRING REVISION OF THE IMPLANT. THE LOOSENING APPEARED RECENTLY AND OTHER THAN THAT THE PATIENT IS DOING WELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
574596 KINOS AXIOM TOTAL ANKLE SYSTEM TOTAL ANKLE HSN RESTOR3D, INC. 1801.1131 2008019

Patients

Seq Age Sex Outcome Treatment
1 NA Male Required Intervention