FDA Adverse Event Malfunction Summary report: N

SUPERDIMENSION INREACH BRONCHUS SYSTEM (SDBS)

MDR report key: 2020507 · Received February 10, 2011

Report

Report Number
3004962788-2011-00007
Event Type
Malfunction
Date Received
February 10, 2011
Date of Event
January 12, 2011
Report Date
February 9, 2011
Manufacturer
SUPERDIMENSION INC.
Product Code
JAK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DE, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

METHOD, RESULTS, AND CONCLUSIONS: A SITE VISIT WAS PERFORMED ON (B)(6) 2011. THE REPORT STATED THAT THERE WAS A FAILURE OF A COMPONENT OF THE SYSTEM, THE LOCATION BOARD CABLE TAIL. THE LOCATION BOARD CABLE TAIL WAS REPLACED AND THE OLD ONE WAS RETURNED FOR ANALYSIS. THE EVALUATION OF THE LOCATION BOARD CABLE TAIL DETERMINED THAT THERE WAS A CABLE CONDUCTOR FRACTURE. THERE WAS NO INJURY TO THE PATIENT REPORTED. IN AN ABUNDANCE OF CAUTION, THIS EVENT IS BEING REPORTED DUE TO THE ADDITIONAL POTENTIAL RISK ASSOCIATED WITH MULTIPLE EXPOSURES TO GENERAL ANESTHESIA.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SITE WAS NOT ABLE TO GET THE LOCATION BOARD TO CONNECT PRIOR TO A PROCEDURE. THEY SWAPPED FOR A SPARE LOCATION BOARD CABLE BUT STILL ENCOUNTERED THE SAME PROBLEM. THEREFORE, THE CASE WAS CANCELLED WITH THE PATIENT UNDER GENERAL ANESTHESIA. THERE WAS NO HARM OR INJURY TO THE PATIENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SUPERDIMENSION INREACH BRONCHUS SYSTEM (SDBS) COMPUTED TOMOGRAPHY X-RAY SYSTEM JAK SUPERDIMENSION INC.