FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2020468 · Received February 16, 2011

Report

Report Number
3004209178-2011-80435
Event Type
Malfunction
Date Received
February 16, 2011
Date of Event
November 25, 2010
Report Date
February 5, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING MANUAL PRIME. TROUBLESHOOTING WAS PERFORMED. RAN A DISPLACEMENT TEST AND THE DEVICE FAILED THE TEST. ADVISED THE CUSTOMER TO RETURN THE INSULIN PUMP AND TO CONTINUE WITH BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.