FDA Adverse Event Malfunction Summary report: N

LC PCA III INFUSER

MDR report key: 2020401 · Received February 8, 2011

Report

Report Number
9615050-2011-00060
Event Type
Malfunction
Date Received
February 8, 2011
Date of Event
January 3, 2011
Report Date
January 10, 2011
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
MEA
PMA / PMN Number
K043256
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT THE DEVICE DID NOT SOUND AN AUDIBLE ALARM TONE DURING AN ALARM CONDITION. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT WITH AN UNSIGNED NOTE THAT STATE, "VERY LOW VOLUME EVEN DURING SOME ALARMS." NO TRACKING INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. DURING TESTING AT THE USER FACILITY, THE DEVICE DID NOT SOUND AN AUDIBLE ALARM TONE DURING AN ALARM CONDITION. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LC PCA III INFUSER 80MEA MEA HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK