FDA Adverse Event Injury Summary report: N

VISCOAT

MDR report key: 2020365 · Received March 11, 2011

Report

Report Number
3002037047-2011-00005
Event Type
Injury
Date Received
March 11, 2011
Date of Event
September 1, 2008
Report Date
February 9, 2011
Manufacturer
ALCON - BELGIUM/ S.A. ALCON-COUVREUR N.V.
Product Code
LZP
PMA / PMN Number
P840064
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT WAS NOT RETURNED FOR ANALYSIS. PRODUCT HISTORY RECORDS COULD NOT BE REVIEWED BECAUSE THE REPORTING FACILITY DID NOT PROVIDE A LOT NUMBER OR ANY IDENTIFICATION TRACEABLE TO THE MFG DOCUMENTATION. LITERATURE CITATION: YUAN-CHIEH LEE: THERMAL BURNS CAUSED BY OPHTHALMIC VISCOSURGICAL DEVICE OCCLUSION IN TORSIONAL PHACOEMULSIFICATION: TZU CHI MEDICAL JOURNAL. DECEMBER 2010, VOLUME 22, NUMBER 4. (B)(4) (DEVICE OPERATIONAL ISSUE). (B)(4).

Description of Event or Problem · 1

LITERATURE ARTICLE REPORTS THREE CASES OF PHACO BURNS DURING TORSIONAL PHACOEMULSIFICATIONS, WHICH WERE, IN THE RPTR'S OPINION, CAUSED BY THE OPHTHALMIC VISCOSURGICAL DEVICE OCCLUSION. IN ALL THREE CASES, THE PHACO BURNS OCCURRED AT THE BEGINNING OF SCULPTING. THE CORNEAL WOUNDS WERE SUTURED, INTENTIONALLY LEAVING THE ADHESIONS OF THE IRIS TO THE CORNEAL WOUNDS. SEPARATION OF THE CORNEA AND IRIS WERE DONE LATER, AND A CENTRAL ROUND PUPIL WAS ACHIEVED IN CASES TWO AND THREE. ASTIGMATISM IMPROVED WITH TIME IN ALL THREE CASES. THIS IS THE SECOND OF THREE MEDICAL DEVICE REPORTS BEING FILED FOR THIS LITERATURE REPORT AND REPRESENTS CASE 2.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VISCOAT AID, SURGICAL, VISCOELASTIC LZP ALCON - BELGIUM/ S.A. ALCON-COUVREUR N.V. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 55 YR Required Intervention INFINITI TORSIONAL PHACO HANDPIECE| INFINITI VISION SYSTEM OZIL| INFINITI VISION SYSTEM PAK