FDA Adverse Event
Malfunction
Summary report: N
AFFINITY BED
MDR report key: 2020133
·
Received March 10, 2011
Report
- Report Number
- 1824206-2011-01421
- Event Type
- Malfunction
- Date Received
- March 10, 2011
- Date of Event
- February 18, 2011
- Report Date
- February 18, 2011
- Manufacturer
- HILL-ROM INC.
- Product Code
- HDD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TECHNICIAN REPORTED THAT WHILE HE WAS PERFORMING AN ELECTRICAL SAFETY CHECK ON THE BED, THAT THE GROUND RESISTANCE WAS TOO HIGH. REPLACED THE POWER CORD TO RESOLVE THIS ISSUE. THE BED WAS LOCATED IN THE LABOR AND DELIVERY WARD.
Description of Event or Problem · 1
INFO RECEIVED INDICATED THAT THE GROUND RESISTANCE WAS TOO HIGH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AFFINITY BED | BIRTHING BED | HDD | HILL-ROM INC. | P3700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |