FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2020132 · Received March 10, 2011

Report

Report Number
1824206-2011-01427
Event Type
Malfunction
Date Received
March 10, 2011
Date of Event
February 19, 2011
Report Date
February 19, 2011
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN FOUND RIBBON CABLE CONNECTING RECEPTACLE TO BIU WAS BROKEN. REPLACED RIBBON CABLE ((B)(4)) TO RESOLVE THIS ISSUE.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT PT IN MEDSURG ROOM CAN PLACE NURSE CALL, WITH RESPONSE NOT HEARD IN EXPECTED LOCATION. NO INJURY ALLEGED BY ACCOUNT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. P1900

Patients

Seq Age Sex Outcome Treatment
1