FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE
MDR report key: 2020132
·
Received March 10, 2011
Report
- Report Number
- 1824206-2011-01427
- Event Type
- Malfunction
- Date Received
- March 10, 2011
- Date of Event
- February 19, 2011
- Report Date
- February 19, 2011
- Manufacturer
- HILL-ROM INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TECHNICIAN FOUND RIBBON CABLE CONNECTING RECEPTACLE TO BIU WAS BROKEN. REPLACED RIBBON CABLE ((B)(4)) TO RESOLVE THIS ISSUE.
Description of Event or Problem · 1
COMPLAINANT ALLEGED THAT PT IN MEDSURG ROOM CAN PLACE NURSE CALL, WITH RESPONSE NOT HEARD IN EXPECTED LOCATION. NO INJURY ALLEGED BY ACCOUNT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM INC. | P1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |