FDA Adverse Event
Malfunction
Summary report: N
CARE ASSIST
MDR report key: 2020115
·
Received March 10, 2011
Report
- Report Number
- 1824206-2011-01424
- Event Type
- Malfunction
- Date Received
- March 10, 2011
- Date of Event
- February 18, 2011
- Report Date
- February 18, 2011
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TECHNICIAN FOUND BED IN THE MAINTENANCE SHOP WITH A NOTE ON IT STATING THE HEAD SECTION OF THE BED WOULD NOT LOCK WHEN THE BRAKE WAS SET. TECHNICIAN FOUND THE BRAKE BAR WAS DISCONNECTED FROM THE HEAD LEFT CASTER. RECONNECTED THE BRAKE BAR TO RESOLVE THIS ISSUE.
Description of Event or Problem · 1
COMPLAINT ALLEGED THAT THE HEAD SECTION OF THE BED DOES NOT LOCK WHEN THE BRAKE WAS SET.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CARE ASSIST | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | P1170 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |