FDA Adverse Event Malfunction Summary report: N

GPS PROCEDURAL STRETCHER

MDR report key: 2020089 · Received March 10, 2011

Report

Report Number
1824206-2011-01413
Event Type
Malfunction
Date Received
March 10, 2011
Date of Event
February 14, 2011
Report Date
February 14, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE TOP OF THE CASTER STEMS ARE BROKEN AND WILL NOT ALLOW THE BRAKE TO BE SET. HE REPLACED THE BRAKE AND BRAKE CASTERS TO REPAIR THE STRETCHER.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE BRAKE AND STEER CASTERS DO NOT HOLD WHEN SET IN BRAKE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GPS PROCEDURAL STRETCHER WHEELED STRETCHER FPO HILL-ROM, INC. 883

Patients

Seq Age Sex Outcome Treatment
1