FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2020069 · Received March 9, 2011

Report

Report Number
3004209178-2011-80590
Event Type
Malfunction
Date Received
March 9, 2011
Date of Event
February 17, 2011
Report Date
February 21, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS NOT RECEIVING ALL THE INSULIN IN HER RESERVOIR AND WAKING WITH HIGH BLOOD GLUCOSE LEVELS. TROUBLESHOOTING WAS PERFORMED. THE INSULIN PUMP PASSED THE FIXED PRIME BUT FAILED THE HIGH PRESSURE TEST AND THE ALARM HISTORY DID NOT SHOW ANY NO DELIVERY ALARMS, ONLY LOW RESERVOIR ALARMS. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAP

Patients

Seq Age Sex Outcome Treatment
1 35 YR