NV GXL LINER LIPPED 36MM ID GROUP 2
Report
- Report Number
- 1038671-2024-03483
- Event Type
- Injury
- Date Received
- September 12, 2024
- Date of Event
- October 18, 2023
- Report Date
- February 7, 2025
- Manufacturer
- EXACTECH, INC.
- Product Code
- JDI
- UDI-DI
- 10885862207036
- PMA / PMN Number
- K121392
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- 003
Narratives
D10: CONCOMITANT DEVICES: (B)(6) 170-36-93 - BIOLOX DELTA FEMORAL HEAD 36MM OD, -3.5MM (B)(6) 188-00-08 - WEDGE PLASMA S/O SZ 8 (B)(6) 164-11-12 - NOVATION ELEMENT RO S/O SZ 12 (B)(6) 186-01-52 - INTEGRIP CC, CLUSTER 52MM, G2 (B)(6) 170-36-00 - BIOLOX DELTA FEMORAL HEAD 36MM OD, +0MM. THE DEVICE WAS NOT RETURNED FOR EVALUATION AND NO MEDICAL OR OTHER RECORDS CONTAINING TREATMENT INFORMATION OR PATIENT INFORMATION HAVE BEEN RECEIVED; THEREFORE, THE REPORTED EVENT CANNOT BE CONFIRMED, NOR CAN THE CIRCUMSTANCES OR POTENTIAL CAUSES OR CONTRIBUTORS TO THE ALLEGED EVENT BE EVALUATED. SHOULD ADDITIONAL, MATERIAL INFORMATION BECOME AVAILABLE THAT PERMITS MORE ANALYSIS OR CONCLUSIONS, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL AND/OR CORRECTED INFORMATION. THE FOLLOWING SECTIONS WERE CORRECTED: H6 (HEALTH EFFECT - CLINICAL CODE). THE MOST LIKELY CAUSE FOR THE REPORTED REVISION DUE TO PROSTHESIS WEAR IS A COMBINATION OF THE RISK FACTORS SPECIFIED SUCH AS USE ERROR, IMPLANT POSITIONING, IMPLANT SIZE SELECTION, AND PATIENT FACTORS (FITNESS FOR SURGERY, BIOMECHANICS, ACTIVITY LEVEL AND LOCAL TISSUE OXIDATION POTENTIAL). HOWEVER, THIS CANNOT BE CONFIRMED AS THE DEVICES WERE NOT RETURNED FOR EVALUATION, AND IMAGES AND RADIOGRAPHS WERE NOT PROVIDED. IF ANY FURTHER INFORMATION IS OBTAINED THAT WOULD CHANGE OR ALTER ANY INFORMATION PROVIDED, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.
IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 58 MONTHS AFTER A LEFT TOTAL HIP REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 557774 | NV GXL LINER LIPPED 36MM ID GROUP 2 | PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED | JDI | EXACTECH, INC. | 10885862207036 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Required Intervention| H | SEE H11 |