VENTED AUTOFEED HUMIDIFICATION CHAMBER
Report
- Report Number
- 9611451-2011-00160
- Event Type
- Malfunction
- Date Received
- March 15, 2011
- Date of Event
- February 21, 2011
- Report Date
- February 21, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). METHOD: THE RETURNED MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER WAS CONNECTED TO A WATERBAG FOR TESTING AND WAS VISUALLY INSPECTED FOR DAMAGE. RESULTS: THE CHAMBER FILLED TO A NORMAL LEVEL DURING TESTING. WHEN THE WATER FLOW STOPPED, SMALL DROPS OF WATER WERE OBSERVED TO LEAK SLOWLY FROM THE CONNECTION BETWEEN THE WATER FEEDSET TUBE AND THE WATERBAG SPIKE. VISUAL INSPECTION SHOWED THAT INSUFFICIENT AMOUNT OF GLUE HAD BEEN APPLIED BETWEEN THE WATER FEEDSET TUBE AND SPIKE, CAUSING THE REPORTED LEAK. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 100326. CONCLUSION: BASED ON VISUAL INSPECTION OF THE FEEDSET TUBE AND WATERBAG SPIKE OF THE RETURNED MR290 CHAMBER, THE LEAK WAS CAUSED BY AN INSUFFICIENT AMOUNT OF GLUE APPLIED. THE USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 CHAMBER STATE "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." AS PART OF OUR ONGOING PRODUCT IMPROVEMENTS, ADDITIONAL GLUE IS NOW APPLIED TO THE FEEDSET SPIKE DURING PRODUCTION. (B)(4).
A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT WATER LEAKED BETWEEN THE WATER FEED TUBE AND SPIKE OF AN MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER. THIS WAS NOTICED BEFORE PATIENT USE. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VENTED AUTOFEED HUMIDIFICATION CHAMBER | BTT | BTT | FISHER & PAYKEL HEALTHCARE LIMITED | MR290V | 100326 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |