FDA Adverse Event Malfunction Summary report: N

ARCHITECT STAT TROPONIN-I REAGENT KIT

MDR report key: 20194024 · Received September 11, 2024

Report

Report Number
1415939-2024-00054
Event Type
Malfunction
Date Received
September 11, 2024
Date of Event
August 22, 2024
Report Date
October 14, 2024
Manufacturer
ABBOTT LABORATORIES
Product Code
MMI
UDI-DI
00380740048525
PMA / PMN Number
K041192
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED FROM ARCHITECT STAT TROPONIN-I REAGENT KIT, LIST NUMBER 02K41-27, ABBOTT PARK, IL, USA IN SECTION D OF THIS REPORT TO ARCHITECT I1000SR, LIST NUMBER 01L86-40, IRVING, TX, USA. MDR NUMBER 3016438761-2024-00603-00 HAS BEEN SUBMITTED AND ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER THAT MDR NUMBER.

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED A FALSELY ELEVATED ARCHITECT STAT TROPONIN-I RESULT FOR ONE PATIENT. THE FOLLOWING DATA WAS PROVIDED (NORMAL RANGE IS 0-29 NG/ML): PATIENT 1 INITIAL RESULT WAS 141.6, REPEAT RESULT WAS <7, REPEAT RESULT, ON ANOTHER ANALYZER, WAS <7 NG/ML. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED A FALSELY ELEVATED ARCHITECT STAT TROPONIN-I RESULT FOR ONE PATIENT. THE FOLLOWING DATA WAS PROVIDED (NORMAL RANGE IS 0-29 NG/ML): PATIENT 1 INITIAL RESULT WAS 141.6, REPEAT RESULT WAS <7, REPEAT RESULT, ON ANOTHER ANALYZER, WAS <7 NG/ML. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
557096 ARCHITECT STAT TROPONIN-I REAGENT KIT IMMUNOASSAY METHOD, TROPONIN SUBUNIT MMI ABBOTT LABORATORIES 07500UN24 00380740048525

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown ARC I1000SR INTGR, 01L86-40, (B)(6).| ARC I1000SR INTGR, 01L86-40, (B)(6).