FDA Adverse Event Malfunction Summary report: N

AUTOMATED IMPELLA CONTROLLER

MDR report key: 20189954 · Received September 11, 2024

Report

Report Number
1220648-2024-17628
Event Type
Malfunction
Date Received
September 11, 2024
Date of Event
August 14, 2024
Report Date
October 4, 2024
Manufacturer
ABIOMED, INC.
Product Code
OZD
UDI-DI
00813502011401
PMA / PMN Number
P140003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE IMPELLA DEVICE WAS RETURNED, AND AN EVALUATION IS UNDERWAY. UPON COMPLETION OF OUR ANALYSIS, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

THE INVESTIGATION FOR THE CONSOLE (AIC) ALARM MALFUNCTIONS AND SHUTDOWN ISSUES HAS BEEN COMPLETED. CONSOLE LOGS FROM THE REPORTED DAY OF EVENT SHOW AN UNEXPECTED SHUTDOWN AND RESTART OF THE CONSOLE, FOLLOWED BY ¿UNEXPECTED SHUTDOWN¿ ALARM AND PRECEDED BY AN ONSET OF PUD COMMUNICATION FAILURES, THAT WERE NOT RESOLVED AFTER PUD RESETS. THE CONSOLE WAS RETURNED FOR EVALUATION FINDING THAT THE ISSUE WAS NOT REPRODUCED. THERE WAS NO NOISE (FROM PBM OR ANYWHERE ELSE), NO UNEXPECTED RESTARTS, OR ALARMS. INSPECTION OF THE WIRING INSIDE THE CONSOLE, HOWEVER, REVEALED LOOSE CONNECTOR OF THE CABLE IN J16 SOCKET ON THE PBM BOARD. THIS CONNECTION SUPPLIES VOLTAGE TO PUD VIA 4-IN-1 CARRIER, VOLTAGE TO EPC, AND GND FOR PUD AND EPC. OBSERVED SYMPTOMS DESCRIBED IN THE COMPLAINT ARE CONSISTENT WITH INTERMITTENT LOSS OF ELECTRICAL CONNECTION AT THIS PORT, AND THE REPORTED NOISE MIGHT HAVE BEEN RELATED TO INTERMITTENT OPERATION OF THE PURGE MOTOR, RATHER THAN PBM BOARD BUZZER. LOOSE CONNECTOR WAS MOST LIKELY RESULT OF AN ERROR DURING ONE OF PRIOR PM OR FC PROCEDURES OF THE LOANER IC9502. ADDED- H6 COMPONENT CODE 4733.

Description of Event or Problem · 0

THE AUTOMATED IMPELLA CONTROLLER (AIC) WAS IN FOR SERVICE AND FOUND TO HAVE A MALFUNCTION. THE AIC WAS NOISY AND KEPT RESTARTING DURING THE CONSOLE CHECK. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1510977 AUTOMATED IMPELLA CONTROLLER CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS OZD ABIOMED, INC. AUTOMATED IMPELLA CONTROLLER 1637920 00813502011401

Patients

Seq Age Sex Outcome Treatment
1 0 DA Unknown