FDA Adverse Event Malfunction Summary report: N

TELESCOPE "OES ELITE", 4 MM, 12°, HD, AUTOCLAVABLE

MDR report key: 20189336 · Received September 11, 2024

Report

Report Number
9610773-2024-02184
Event Type
Malfunction
Date Received
September 11, 2024
Date of Event
September 4, 2024
Report Date
November 15, 2024
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
FGC
UDI-DI
04042761077507
PMA / PMN Number
K151191
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION, AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE RESULTS OF THE INVESTIGATION, THE LIKELY CAUSE OF THE REPORTED EVENT IS DUE TO IMPROPER HANDLING AND APPLICATION OF EXCESSIVE FORCE, IMPACT TO THE DEVICE LIKE FALL, SHOCK OR SIMILAR STRESS. HOWEVER, THE ROOT CAUSE OF THE REPORTED EVENT IS UNABLE TO BE DETERMINED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Additional Manufacturer Narrative · 0

THE EVALUATION OF THE EVENT IS ONGOING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THE TELESCOPE HAD A LOOSE BLUE RING COMPONENT. THERE WERE NO REPORTS OF PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
567203 TELESCOPE "OES ELITE", 4 MM, 12°, HD, AUTOCLAVABLE WA2T412A FGC OLYMPUS WINTER & IBE GMBH WA2T412A 803508 04042761077507

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown