FDA Adverse Event Injury Summary report: N

LOGIC TIBIA IMPLANT PSC INSERT, SZ 1, 13MM

MDR report key: 20189100 · Received September 11, 2024

Report

Report Number
1038671-2024-03436
Event Type
Injury
Date Received
September 11, 2024
Date of Event
August 29, 2023
Report Date
October 22, 2024
Manufacturer
EXACTECH, INC.
Product Code
JWH
UDI-DI
10885862173416
PMA / PMN Number
K110547
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE PRODUCT ASSOCIATED WITH THE REPORTED EVENT IS WITHIN THE SCOPE OF RECALL Z-0021-2022. HOWEVER, THERE IS INSUFFICIENT INFORMATION TO EVALUATE WHETHER THE SUBJECT ISSUE OF THE RECALL WAS A CAUSE OR CONTRIBUTOR TO THE REPORTED EVENT. THE DEVICE WAS NOT RETURNED FOR EVALUATION AND NO MEDICAL OR OTHER RECORDS CONTAINING TREATMENT INFORMATION OR PATIENT INFORMATION HAVE BEEN RECEIVED; THEREFORE, THE REPORTED EVENT CANNOT BE CONFIRMED, NOR CAN THE CIRCUMSTANCES OR POTENTIAL CAUSES OR CONTRIBUTORS TO THE ALLEGED EVENT BE EVALUATED. SHOULD ADDITIONAL, MATERIAL INFORMATION BECOME AVAILABLE THAT PERMITS MORE ANALYSIS OR CONCLUSIONS, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Additional Manufacturer Narrative · 0

H6: CORRECTED THE FOLLOWING: HEALTH EFFECT - CLINICAL CODE, TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. THE REASON THE REVISION REPORTED CANNOT BE CONFIRMED FROM THE INFORMATION PROVIDED BUT MAY BE THE RESULT OF PROSTHESIS WEAR AND LOOSENING OR DUE TO INCLUSION OF THE POLYETHYLENE IN THE PACKAGING RECALL. POTENTIAL CONTRIBUTIONS OF USER AND PATIENT-RELATED CONSIDERATIONS TO THE EVENT COULD NOT BE ASSESSED AS THE DEVICES WERE NOT AVAILABLE FOR EVALUATION AND IMAGES, RADIOGRAPHS, AND RELEVANT CLINICAL INFORMATION WERE NOT PROVIDED.

Description of Event or Problem · 0

LEGAL CASE ¿ (B)(4) LONG FORM IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 76 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT HAS EXPERIENCED PROSTHESIS WEARPAIN, DISCOMFORT, GAIT IMPAIRMENT, BONE LOSS. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE. (B)(6) LOGIC TIBIA IMPLANT PSC SERIAL NUMBER (B)(6) IS CONFIRMED TO HAVE BEEN PACKAGED IN A VACUUM BAG THAT DOES NOT CONTAIN EVOH. 510K: K110547.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
612268 LOGIC TIBIA IMPLANT PSC INSERT, SZ 1, 13MM PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL JWH EXACTECH, INC. UNK 10885862173416

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention| H