FDA Adverse Event Malfunction Summary report: N

RAYONE EMV

MDR report key: 20189091 · Received September 11, 2024

Report

Report Number
3012304651-2024-00231
Event Type
Malfunction
Date Received
September 11, 2024
Report Date
September 11, 2024
Manufacturer
RAYNER INTRAOCULAR LENSES LIMITED
Product Code
HQL
PMA / PMN Number
PO60011
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE REFERENCE (B)(4) HAS BEEN ALLOCATED TO THIS CASE BY RAYNER. THE EVENT DESCRIPTION PROVIDED STATES THAT POST-OPERATIVELY, THE PATIENT PRESENTED WITH TOXIC ANTERIOR SEGMENT SYNDROME (TASS). THE AETIOLOGY OF TASS MAY BE MULTI-FACTORIAL WITH NUMEROUS POTENTIAL CAUSES INCLUDING BUT NOT LIMITED TO; BACTERIAL ENDOTOXINS OR PARTICULATE CONTAMINATION OF BALANCED SALT SOLUTIONS, INTRAOCULAR IRRIGATING SOLUTIONS WITH ABNORMAL PH, OSMOLARITY OR IONIC COMPOSITION, DENATURED OPHTHALMIC VISCOSURGICAL DEVICES (OVD), INTRAOCULAR MEDICATIONS (ANTIBIOTICS IN THE IRRIGATION SOLUTIONS OR INTRACAMERAL ANTIBIOTICS),TOPICAL OINTMENTS, INADEQUATE STERILIZATION OF SURGICAL INSTRUMENTS AND TUBING, INADEQUATE FLUSHING OF INSTRUMENTS BETWEEN CASES RESULTING IN BUILD-UP OF OPHTHALMIC VISCOSURGICAL DEVICES (OVD),PRESERVATIVE AND METALLIC PRECIPITATE. EVERY BATCH OF IOLS MANUFACTURED BY RAYNER ARE TESTED FOR ENDOTOXIN AND BIOBURDEN PRIOR TO RELEASE FOR SALE. RAYNER'S ENDOTOXIN ACCEPTABLE TOLERANCES ARE DERIVED FROM BS EN ISO11979-08 (2017) OPHTHALMIC IMPLANTS - INTRAOCULAR LENSES PART 8 FUNDAMENTAL REQUIREMENTS. WE HAVE SET OUR OWN LIMITS FOR BIOBURDEN AS THERE IS NO STANDARD THAT PRESCRIBES ACCEPTABLE LIMITS. I CAN CONFIRM THAT THE RECORDS FOR THE RAYONE EMV RAO200E BATCH 063217851 SHOW THAT THE BIOBURDEN AND ENDOTOXIN RESULTS ARE WITHIN THE DEFINED LIMITS. A RAYNER MICROBIOLOGICAL CAUSE FOR THE ONSET OF POST-OPERATIVE TASS IS THEREFORE EXTREMELY UNLIKELY. RAYNER HAS BEEN ADVISED THAT THE HEALTHCARE FACILITY HAS COMPLETED THEIR OWN INVESTIGATION AND HAS ALSO CONCLUDED THAT THERE IS NO CAUSAL RELATION BETWEEN THE ONSET OF TASS AND THE IMPLANTED RAYONE EMV RAO200E IOL.

Description of Event or Problem · 0

ON 22ND AUGUST 2024, RAYNER RECEIVED NOTIFICATION FROM A US HEALTHCARE FACILTIY OG AN EVENT THAT OCCURRED FOLLOWING IMPLANTATION OF A RAYONE EMV RAO200E. THE EVENT DESCRIPTION PROVIDED STATES THAT POST-OPERATIVELY, THE PATIENT PRESENTED WITH TOXIC ANTERIOR SEGMENT SYNDROME (TASS).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
612259 RAYONE EMV RAYONE EMV HQL RAYNER INTRAOCULAR LENSES LIMITED RAO200E 063217851

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown