FDA Adverse Event
Injury
Summary report: N
IMPLANTABLE COLLAMER LENS (ICL)
MDR report key: 20186176
·
Received September 10, 2024
Report
- Report Number
- 2023826-2024-04200
- Event Type
- Injury
- Date Received
- September 10, 2024
- Date of Event
- April 7, 2024
- Report Date
- August 16, 2024
- Manufacturer
- STAAR SURGICAL
- Product Code
- QCB
- PMA / PMN Number
- P030016
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
MANUFACTURER NARRATIVE - SECONDARY SURGICAL INTERVENTION TO REMOVE/REPLACE THE LENS IS IDENTIFIED IN THE LABELING AS A KNOWN POTENTIAL ADVERSE EVENT FOLLOWING ICL IMPLANTATION. (B)(4).
Additional Manufacturer Narrative · 0
CORRECTION: G4- PREMARKET IDENTIFICATION: DEVICE BLA: P030016 TO PMA/510(K): P030016, INITIALLY REPORTED PMA/510(K) NUMBER IN THE WRONG FIELD. (B)(4).
Description of Event or Problem · 0
A HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN IMPLANTABLE COLLAMER LENS (ICL) IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED LOW VAULT WITHOUT ROTATION. IN A SEPARATE VISIT THE LENS WAS EXCHANGED FOR A LONGER LENGTH LENS AND THE PROBLEM RESOLVED. CAUSE OF THE EVENT WAS REPORTS AS UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 562976 | IMPLANTABLE COLLAMER LENS (ICL) | PHAKIC TORIC INTRAOCULAR LENS | QCB | STAAR SURGICAL | VTICMO12.1 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR | Female | Required Intervention |