FDA Adverse Event Injury Summary report: N

STYLE 10 SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 20182133 · Received September 10, 2024

Report

Report Number
9617229-2024-20375
Event Type
Injury
Date Received
September 10, 2024
Report Date
November 25, 2024
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.

Additional Manufacturer Narrative · 0

ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: B.5 ,D.6A, D.6B, H.6.

Additional Manufacturer Narrative · 0

LABORATORY ANALYSIS SUMMARY: THE DEVICE RELATED TO THE REPORTED EVENTS RUPTURE WAS RECEIVED ON OCTOBER 11, 2024, WITH LOT NUMBER 1163096. BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS ARE: ¿ RUPTURE: OBSERVED OPENING ON THE RADIUS SIDE ASSESSED AS FOLD FLAW OPENING. AS PER THE INVESTIGATION PROCEDURE, NON-PENETRATING NICKS, WEAR ABRASION AND CREASES WERE COMPLETED AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "RUPTURE." RECORD IS FOR RIGHT SIDE. DEVICE REMAINS IMPLANTED.

Description of Event or Problem · 0

DEVICE HAS BEEN EXPLANTED.

Description of Event or Problem · 0

A "POPCORN CAPSULORRHAPHY" WAS PERFORMED DURING EXPLANT SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2298008 STYLE 10 SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 1163096

Patients

Seq Age Sex Outcome Treatment
1 NA Female Required Intervention