CORDELLA¿ PULMONARY ARTERY SENSOR SYSTEM
Report
- Report Number
- 3024985933-2024-00008
- Event Type
- Injury
- Date Received
- September 10, 2024
- Date of Event
- July 26, 2024
- Report Date
- May 15, 2025
- Manufacturer
- ENDOTRONIX, INC
- Product Code
- MOM
- UDI-DI
- 00850008997006
- PMA / PMN Number
- P230040
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
CORRECTED DATA: SECTION G3: INITIAL REPORT DATE RECEIVED BY MANUFACTURER.
UPDATED SECTION(S): G6 AND H3: SECTION B5: ADDITIONAL INFORMATION RECEIVED INDICATES THAT A RIGHT HEART CATHETERIZATION (RHC) WAS PERFORMED TO CONFIRM THE ACCURACY OF THE PRESSURE SENSOR AGAINST A REFERENCE PRESSURE. THE SENSOR WAS RECALIBRATED, AND AN ADJUSTMENT WAS REQUIRED. SECTION B6: (B)(6) 2024 - CALIBRATION DURING RHC VIA SWAN-GANZ CONFIRMED THAT A SENSOR ADJUSTMENT OF 17.386MMHG WAS REQUIRED. ADDITIONAL INFORMATION WILL BE SUBMITTED VIA FOLLOW-UP REPORT WITHIN 30 DAYS OF RECEIPT.
(H3) THE SENSOR WAS NOT RETURNED FOR EVALUATION AS IT REMAINS IMPLANTED. THE DEVICE HISTORY RECORD (DHR) FOR THE SENSOR LOT WAS REVIEWED AND IT WAS CONFIRMED THAT ALL MANUFACTURING OPERATIONS AND INSPECTIONS WERE PERFORMED, ALL ACCEPTANCE CRITERIA WERE MET, AND NO RELEVANT NONCONFORMANCES WERE ASSOCIATED WITH THE SENSOR LOT AT THE TIME OF MANUFACTURE. ON (B)(6) 2024, THE PATIENT UNDERWENT A RIGHT HEART CATHETERIZATION (RHC) RECALIBRATION. THE OFFSET DETERMINED DURING THE RECALIBRATION FELL OUTSIDE THE FLUID-FILLED REFERENCE MEASUREMENT ERROR RANGE, THEREBY CONFIRMING THAT THE SENSOR WAS PROVIDING INACCURATE MEASUREMENTS. FOLLOWING RECALIBRATION, THE MPAP TREND HAS REMAINED STABLE AND CONSISTENT. WHILE THE EXACT CAUSE OF THE REPORTED EVENT CANNOT BE CONCLUSIVELY DETERMINED, THE MEASUREMENT STABILITY AFTERWARDS INDICATES THAT THE SENSOR'S INACCURATE MEASUREMENT WAS FINITE, AND HAS BEEN EFFECTIVELY CORRECTED THROUGH RECALIBRATION.
ADDITIONAL INFORMATION WILL BE SUBMITTED VIA FOLLOW-UP REPORT WITHIN 30 DAYS OF RECEIPT.
SENSOR INACCURACY SUSPECTED. PATIENT WILL UNDERGO REGULARLY SCHEDULED RECALIBRATION ON 26-SEPT-2024 TO DETERMINE IF AN ADJUSTMENT IS NEEDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 221701 | CORDELLA¿ PULMONARY ARTERY SENSOR SYSTEM | PULMONARY ARTERY SENSOR | MOM | ENDOTRONIX, INC | 00850008997006 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Male | Required Intervention |