ARIA
Report
- Report Number
- 2916710-2011-00021
- Event Type
- Malfunction
- Date Received
- March 8, 2011
- Date of Event
- August 6, 2010
- Report Date
- February 8, 2011
- Manufacturer
- VARIAN MEDICAL SYSTEMS INC.
- Product Code
- IYE
- PMA / PMN Number
- K093527
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIST
Narratives
THE COMBINATION OF ARIA 8.9 AND OBI 1.4 CAN LEAD TO THE DRR, KV AND MV IMAGES APPEARING AS COMPLETELY SHIFTED AND MISMATCHED, WHEN LOOKING AT THE ON-LINE MATCHING RESULT, IN OFF-LINE REVIEW. THE IMAGES APPEAR AS IF THEY WERE NOT MATCHED AT ALL, WHEN IN FACT MATCHING HAS BEEN DONE AND SAVED ON-LINE. THIS DISPLAY ISSUE IN OFF-LINE REVIEW CAN OCCUR WHEN THE REFERENCE IMAGES ARE IMPORTED FROM A THIRD PARTY TREATMENT PLANNING SYSTEM, WHEN THE PLAN AND IMAGES DO NOT CONTAIN PATIENT POSITION INFORMATION. THE REFERENCE IMAGES HAVE ISOCENTER INFORMATION, BUT THIS INFORMATION IS NOT EXPORTED DUE TO THE MISSING PATIENT ORIENTATION IN THE IMAGE SERIES. THE DISPLAY OF THE IMAGE IN OFF-LINE REVIEW CAN MISLEAD THE USER AND UNNECESSARY SHIFTS FROM THE CORRECT ISOCENTER POSITION CAN OCCUR. THE INCORRECT SHIFT WHICH IS DISPLAYED IN OFFLINE REVIEW OCCURS IF FOLLOWING CRITERIA IS MET: IMPORTED PLAN DOES NOT CONTAIN PATIENT POSITION, AND IMPORTED REFERENCE IMAGES DO NOT CONTAIN PATIENT POSITION, BUT CONTAIN ISOCENTER INFORMATION OTHER THAN 0.0/0.0/0.0, AND DICOM DATA IS IMPORTED INTO ARIA 8.9, AND IMAGES ARE SET AS REFERENCE IMAGES BEFORE DEFINING THE PATIENT ORIENTATION IN THE PLAN, AND NO PLAN REVISION OR PLAN COPY IS DONE, AND OBI 1.4 OR OBI 1.5 WITH WRONG IMPLEMENTATION UID CONFIGURATION IS USED. THOUGH PROBABILITY OF HARM RESULTING FROM THIS ISSUE HAS BEEN DETERMINED TO BE REMOTE, A CUSTOMER TECHNICAL BULLETIN (CTB) WILL BE SENT TO ALL AFFECTED CUSTOMERS. ALL CORRECTIVE ACTIONS WILL BE HANDLED VIA VARIAN'S CAPA PROCESS. NO ADDITIONAL FOLLOW-UP TO THIS MDR IS EXPECTED.
PATIENT IS LOADED ON THE OBI WORKSTATION. ON-LINE MATCHING IS DONE WITHOUT ANY PROBLEM. WHEN OPENING THE PATIENT IN OLR (OFF-LINE REVIEW), THE DRR AND KV IMAGES APPEAR AS COMPLETELY SHIFTED; WHEN LOOKING AT THE ON-LINE MATCHING RESULT, THE IMAGES APPEAR AS IF THEY WERE NOT MATCHED AT ALL WHEREAS MATCHING HAS BEEN DONE AND SAVED ON-LINE. THE CUSTOMER IMPORTS THE PLAN FROM ISOGRAY TPS TO ARIA AND FROM THAT PLAN, MAKES 2 PLANS: A COPY FOR IMAGING AT THE FIRST SESSION AND THE ORIGINAL PLAN FOR THE TREATMENT. IT WAS FOUND THAT THE IMAGING PLAN IS ALWAYS OK (THE ONE THAT HAS BEEN COPIED) BUT THE PLAN WITH THE TREATMENT FIELDS GIVES THE PROBLEM. AT THE OBI, NO PROBLEM, THE IMAGES DO NOT APPEAR SHIFTED; THE PROBLEM ONLY OCCURS IN OFFLINE REVIEW AND FOR ALL THE SESSIONS (SAME SHIFT BETWEEN REFERENCE IMAGE AND KV IMAGE). IF YOU COPY-PASTE THE PLAN OR DO A PLAN REVISION, THE PROBLEM IS SOLVED. THERE WAS NO INJURY OR MISADMINISTRATION REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ARIA | ACCELERATOR, LINEAR, MEDICAL | IYE | VARIAN MEDICAL SYSTEMS INC. | HRO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |