FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE

MDR report key: 20180275 · Received September 10, 2024

Report

Report Number
1024879-2024-00797
Event Type
Malfunction
Date Received
September 10, 2024
Date of Event
August 12, 2024
Report Date
August 26, 2024
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
FMI
UDI-DI
30382903686071
PMA / PMN Number
K982541
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: "MATERIAL #: 368607. LOT/BATCH #: 3193563. BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL TESTING, EACH HAVING THEIR SAFETY SHIELD ENGAGED, AND NO ISSUES WERE OBSERVED RELATING TO DEFECTIVE LOCKING MECHANISM AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE DEFECTIVE LOCKING MECHANISM. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS."

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING A BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE, THE SAFETY SHIELD BROKE OFF WHEN TRYING TO CLOSE IT OVER THE IV CANNULA. THERE WAS NO REPORT OF ADVERSE IMPACT TO PATIENTS OR USERS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
223523 BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON, DICKINSON & CO., (BD) 3193563 30382903686071

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown